The following problems that occur during the operation of the electromagnetic flowmeter can be handled in the following way

Fault phenomenon of electromagnetic flowmeter: The data is zero when the pipe is full of flow

Rated 4.6/5 based on 432 customer reviews
Share:
Content

Cause of malfunction

1) Check if there are any blockages or poor flow areas in the pipeline.

2) The actual flow rate is relatively small.

3) The flowmeter itself is faulty or the parameter Settings are incorrect

Solution

1) Check the pipeline valves or filter plugs to see if they are clogged or not opened. Clean and open them accordingly.

2) Confirm the actual flow rate. None of the flowmeters can start measuring from 0; they all have a smaller initial flow rate value. The initial flow rate value for DN25 is 0.2m ³ h. The initial flow rate value of DN32 is 0.3m ³ /h.

3) The parameter Settings of the instrument should be checked

The “small-signal cut-off value” is generally recommended at the factory at 0.5.

The “instrument range” is no more than 10 for DN25 and no more than 16 for DN32.

When the electromagnetic flowmeter is in an empty pipe or with fluid flow in the pipeline, observe the change in the size of the FS value during the “flow zero correction”. Generally, the FS value of a normal meter head is very small when the pipe is empty or the fluid is not flowing. As the fluid flows, the FS value will increase (regardless of whether it is positive or negative, as long as there is a change), and it will fluctuate up and down but will not exceed 5.

If the start and stop values of the pump remain unchanged or are 0, it indicates that there is a problem with the meter head and it needs to be returned to the factory for repair. During the debugging process, it is recommended to set “Reverse output Allowed” to “Allowed”.

Inquiry


    More Electromagnetic Flow Meter

    electromagnetic flowmeters have emerged as indispensable tools for industries that demand accurate and reliable flow measurements. Their advanced technology, coupled with the ability to operate in diverse environments and handle various conductive liquids, positions them as a preferred choice in applications ranging from water management to chemical processing. As technology continues to evolve, electromagnetic flowmeters are likely to witness further advancements, contributing to increased efficiency and precision in liquid flow measurement across different sectors.

    Size: DN3-DN3000mm
    Nominal Pressure: 0.6-1.6Mpa(2.5Mpa/4.0Mpa/6.4Mpa… Max 42Mpa)
    Accuracy: +/-0.5%(Standard) +/-0.3% or +/-0.2%(Optional)
    Liner: PTFE, Neoprene, Hard Rubber, EPDM, FEP, Polyurethane, PFA
    Electrode: SUS316L, Hastelloy B, Hastelloy C Titanium, Tantalum, Platinium-iridium
    Structure Type: Integral type, remote type, submersible type, ex-proof type
    Medium Temperature: -20~+60 degC(Integral type)
    Remote type ( Neoprene ,Hard Rubber ,Polyurethane ,EPDM) -10~+80degC
    Remote type(PTFE/PFA/FEP) -10~+160degC
    Ambient Temperature: -20~+60deg C
    Ambient Humidity: 5-100%RH(relative humidity)
    Measuring Range: Max 15m/s
    Conductivity: >5us/cm
    Protection Class: IP65(Standard); IP68(Optional for remote type)
    Process Connection: Flange (Standard), Wafer, Thread, Tri-clamp etc (Optional)
    Output Signal: 4-20mA/Pulse
    Communication: RS485(Standard), HART(Optional),GPRS/GSM (Optional)
    Power Supply: AC220V (can be used for AC85-250V)
    DC24V (can be used for DC20-36V)
    DC12V (optional), Battery Powered 3.6V (optional)
    Power Consumption: <20W Alarm: Upper Limit Alarm / Lower Limit Alarm Self-diagnosis: Empty Pipe Alarm, Exciting Alarm Explosion Proof: ATEX